5.3. Description of the name plate
Chlorine trace 8232
CN1H-A24p
0...+2 V 0...2 ppm
max 40 °C max 0,5 bar
S/N 1234 42/2016
ID: 00565164
Made in Germany
2
3
5
7
4
8
1
6
9
10
1. Name and type of the sensor
2. Manufacturing code
3. Measuring range at nominal slope
4. Maximum medium pressure
5. Manufacturing date
6. Conformity marking
7. Order code
8. Serial number
9. Maximum medium temperature
10. Type of output signal
Fig. 2 : Name plate (example)
6. TECHNICAL DATA
6.1. Operating conditions
Ambient temperature For "trace" chlorine sensor (order code
565164): 0...+40 °C
For chlorine sensor (order codes
568524
1)
or 568523
2)
): 0...+45 °C
Air humidity < 90%, without condensation
Height above sea level Max. 2000 m
1) or other chlorine sensor for operation at constant pH
2) or other chlorine sensor having a reduced pH-dependency
6.2. Conformity to standards and
directives
The applied standards, which verify conformity with the EU Direc-
tives, can be found on the EU Type Examination Certificate and/or
the EU Declaration of Conformity (if applicable).