Code Reasons Solutions
BATTERY OVER
TEMP
14
Battery has overheated
during charge. Station goes
to Resting 00:00 (code 19)
until the battery cools.
Reduce the Charge rate in the
C-code.
If using a non-OEM battery,
the battery thermistor may be
incorrect for the adapter.
Contact Cadex to upgrade the
adapter.
BATTERY
OVERCHARGED
176
Battery SoC (State-of-
charge) is too high for
QuickTest or OhmTest
The battery is discharged
for five minutes, after which
QuickTest restarts.
Discharge is repeated up to
5 times until SoC is below
90%.
If OhmTest is running, the
program is aborted. Discharge
the battery before running
OhmTest again.
Turn off the automatic
OhmTest. If error continues,
battery cannot be run on
QuickTest or OhmTest.
BATTERY
OVERCHARGED
178
Current program completed;
battery SoC (State-of-
charge) is too high.
Discharge the battery for 10
minutes and then charge
again.
Verify that correct contacts
are used.
Battery may have a high
internal resistance. Verify with
OhmTest and discard the
battery.
BATTERY
REMOVED
17
Battery was removed when
the program completed.
This message appears briefly
when the battery is removed.
If message persists, press
ESC.
131