Detailed
Message
Code Reasons Solutions
BATTERY TOO
HOT
13
Battery temperature is too
hot when inserted. The
battery automatically starts
processing when
temperature drops 5°C
below maximum value set in
the C-code Temperature
Sensing.
Allow the battery to cool or
use the analyzer in a cooler
room.
If the battery is non-OEM, the
thermistor may not match that
on the adapter.
BATTERY
UNDERCHARGE
D
175
Battery SoC (State-of-
charge) is too low for
QuickTest or OhmTest
The battery is charged for
five minutes, after which
QuickTest restarts. Charge
is repeated up to 5 times
until SoC is above 20%.
If OhmTest is running, the
program is aborted. Partially
charge the battery before
running OhmTest again.
BATTERY
UNDERCHARGE
D
177
Current program completed;
battery SoC (State-of-
charge) low.
Charge the battery.
CALIBRATION
FAULT
170
Current cannot pass
through shorted terminal
during calibration.
A minimum AWG 10 (2.5 mm)
wire is recommended to short
contacts/leads.
The wires may have
disconnected during the
calibration process.
Repeat calibration procedure.
CAP. IMPROVED
TO TARGET
195
Target capacity is achieved
or exceeded after an initial
warning
See code 115, Target
capacity not met.
133