21
% SETTING
This balance serves percent (%) display by setting the reference sample to 100%.
Step
(1) Register the % unit. (Refer to “8. Registration, cancel, and change of
unit”. When % unit has been already registered, it is not necessary to
register again.
(2) Place the tare on the pan and press the TARE key.
(3) Load the reference sample.
(4) Continuously press the UNIT key to display “
”.
(5) After the stability mark is lit, press the TARE key.
(6) “
” is displayed for several seconds and the balance enters
the % unit display. Minimum displayed value changes as follows
depending on reference sample weight (REF.)
If the % conversion is not possible, “
” is displayed for
several seconds and the balance returns to weight display.
H Type
REF. < Minimum displayed value × 100 % conversion impossible
Minimum displayed value × 100
< REF. < Minimum displayed value × 1000
100%
Minimum displayed value × 1000
< REF. < Minimum displayed value × 10000
100.0%
Minimum displayed value × 10000
< REF. < Minimum displayed value × 100000
100.00%
Minimum displayed value × 100000 < REF.
100.000%
S Type
REF. < Minimum displayed value × 100 % conversion impossible
Minimum displayed value × 100
< REF. < Minimum displayed value × 1000
100%
Minimum displayed value × 1000
< REF. < Minimum displayed value × 10000
100.0%
Minimum displayed value × 10000 < REF.
100.00%