4. Command Code
Described in this section are the command codes which can be used when your CBL balance is
connected to a computer via the IFB-102A RS-232C interface.
CAUTION : The use of characters other than those described here will cause errors in weighing and
data transfer procedures. If an improper code is mistakenly entered, disconnect the balance power cable
for 10 seconds, then reconnect.
Command
code
Function Description
T Taring
Equivalent to the TARE key
D05 Print (output once)
Equivalent to the PRINT key
D06 Autoprint Refer to “16.1 The EP-60A Electronic Printer”.
D01 Continuous output
Continuous output of data in the balance at every
approx. 100 ms.
For less than 1200 bps, it becomes approx. 150 ms.
D09 Output stop Autoprint or continuous output is canceled.
D07
Single output with stability
Information
Printing is made once with stability information.
D03
Continuous output with
stability information
Continuous printing is made with stability
information.
Q ON/OFF selection
Toggles between standby state and measurement
state.
{
}
Echo back
Characters from these command codes until
delimiter are received and transmitted by every
character.
By use this command, message of personal
computer is able to print out EP-60A.
Characters length is under 16 characters, including
delimiter.