Eccentric error
(1) Prepare a sample which weights
approximately ¼of the balance capacity and
move it on the pan in the order as shown right.
Record the results of X1 to X5 in this order.
(2) If the difference (eccentric error) between
readings at the center position and the off center
positions is as follows:
Pan Position of weight
MAINTENANCE
When fouled
When the balance becomes dirty, wipe it off using a soft cloth with a small amount of mild
detergent.
Organic solvents or chemical dusters should not be used as they may damage painted surfaces
and the display panel.
When the balance is placed in a dusty or dirty environment, use the protection cover of standard
accessory.
The pan can be washed with water. Dry the pan sufficiently and mount to the balance.
320H, 3200H, 3200HL 7 Counts
H Series 4 Counts
S Series 2 Counts