TROUBLESHOOTING
For countermeasures having an asterisk, contact the nearest CAS sales or service center.
When What trouble
Cause -> Countermeasure
Before
Weighing
Nothing is displayed by
connecting
the AC adapter to the
outlet.
is displayed.
The AC adapter is disconnected.
The electrical board is turned OFF.
There is an internal error in the
balance. *
During
Weighing
is displayed.
is displayed.
The display fluctuates.
The display does not change
from zero even if a sample
having the weight near readability
is loaded.
The display slowly changes
when
small amount of sample is
loaded.
(Normal: readability/1 sec.)
has appeared
suddenly.
Data communication cannot
be made.
The mass on the pan is too heavy.
Sensitivity is not correct.
The pan or the pan supporter is not in place.
Influence from vibration or wind
Improve the installation site.
Change the measurement mode
to High-stability mode.
Influence from electric noise or
electromagnetic wave
Maintain a proper distance from the
noise source.
Zero tracking works.
Refer to “6. Menu Selection”.
The averaging processing is in
High-stability mode.
Change the measurement mode if necessary.
There has been an instantaneous power failure.
Press the POWER/BRK key (the
balance enters weight display mode).
Setting of communication parameter is wrong.
Refer to “16.4 Setting the input/output format”
Wiring of RS-232C cable is wrong.
During
PCS
or %
setting
has displayed.
Set value exceeds the specified range.
"
Refer to “9. % setting” and “10. PCS setting”.
During
sensitivity
calibration
Does not proceed to next step.
(The stability mark does not light)
has displayed.
Influence from vibration or wind
Improve the installation site.
Change the measurement mode to
High- stability mode.
The weight used for sensitivity calibration is
wrong.
Check the weight and retry the sensitivity
calibration.
There is an internal error in the balance. * 