43
Note 1. Command Mode Table
Data Request Signal of CI-200
0 1 2 3 4 5 6 7 8 9 10 11
Descriptions on
Request Signal
CI-200
Output Signal
D dd K Z CR LF Zero Point Key Received Data Return
D dd K T CR LF Zero Point Key Received Data Return
D dd K G CR LF Gross Weight Key Received Data Return
D dd K N CR LF Net Weight Key Received Data Return
D dd H D CR LF Hold Key Received Data Return
D dd K B CR LF Print Key Received Data Return
D dd K C CR LF Total Print Key Received Data Return
D dd K W CR LF Weight Data Request Signal Received Data Return
D dd I D 0 0 0 0 0 CR LF Device Number Received Data Return
D dd H Y 0 0 0 0 0 CR LF Key Tare Value Received Data Return
D dd H I 0 0 0 0 0 CR LF High Limit(LCD Only) Received Data Return
D dd H L 0 0 0 0 0 CR LF Low Limit (LCD Only) Received Data Return
Note 1. (D : 0x44, dd:00~99, K:0x4B , Z:0x5A , CR : 0×0D, LF: 0×0A)
dd = Device Number (2byte), CR = 0×0D, LF: 0×0A
Ex) If a device number is 10, dd becomes 0x31 and 0x30.
Ex) If you want to operate the zero point key when a device number is 11, the indicator operates zeroing if the hex
code of “ 44 31 31 4B 5A 0D 0A” is sent.