CEL-712 Microdust Pro Real-time Dust Monitor
60 Casella Measurement
Command Z – Stop real-time output
The instrument stops transmitting the real-time data output.
Command I – Interrogate target
The instrument transmits a comma-delimited string of characters that contain
two ID strings.
The first ID string is the number of records stored in the logger.
The second ID string is the firmware version of the instrument.
Example
CasellaCEL,(01234) 844100,K, 11,80-176087AXX<CR><LF>
Command T – Download records stored
The instrument downloads all records stored in its internal memory to the PC.
Example
Log close, 28/07/03,12:13:48, 129.396,
129.445,28/07/03,12:13:42,0 [CR][LF]
28/07/03,12:13:48, 129.361, 1 [CR][LF]
28/07/03,12:13:46, 129.387, 2 [CR][LF]
28/07/03,12:13:44, 129.439, 3 [CR][LF]
Log close, 28/07/03,12:11:30, 240.920,
240.947,28/07/03,12:11:22,4 [CR][LF]
28/07/03,12:11:30, 240.882, 5 [CR][LF]
28/07/03,12:11:28, 240.909, 6 [CR][LF]
28/07/03,12:11:26, 240.897, 7 [CR][LF]
28/07/03,12:11:24, 240.920, 8 [CR][LF]
28/07/03,12:11:22, 240.937, 9 [CR][LF]
28/07/03,12:11:20, 240.935, 10 [CR][LF]
28/07/03,12:11:18, 240.943, 11 [CR][LF]
28/07/03,12:11:16, 240.935, 12 [CR][LF]
Command R – Set instrument ID, time and date
This command enables you to set the time, date and two ID strings remotely.
This information is displayed on the instrument’s status screen.
CAUTION
All records stored in the instrument are erased when you use this command.