6
Calibration
NOTE
The autosampler fork, located in the load position for safe shipping will have to be moved from the
chamber prior to homing the autosampler.
1. Move the autosampler fork to a safe location.
1.1 Select the menu icon at the lower right hand side of the touchscreen.
1.2 From the menu select “Tools” and scroll down to the Autosampler section at the bottom of the
page.
WARNING
Do not ever move hands, or any object, past the automated arms into the area containing the
rack while the EDGE is running a method or the automation is being controlled via tools. Only
load or unload rack when the EDGE is in the idle state.