39
Routine Maintenance
Routine Maintenance
CEM suggests that you perform the routine maintenance on your instrument as outlined below, along with having a
CEM Certied Technician to come check out your system once a year.
WARNING
Proper precautions must be taken to avoid contact with reagents or reagent vapors. Protective gear should
be worn as outlined in the user’s safety program for hazardous materials and the reagent manufacturer’s
safety data sheet. Refer to these guidelines for proper handling and disposal of the reagents. Dispose of
all waste in accordance with all applicable local, state, and federal health and safety recommendations.
After Every use
• Wipe inside of MARS 6 cavity, including temperature sensors, with a lint free damp cloth.
CAUTION
Only use water and a soft cloth to clean the system. Using other cleaning solutions, brushes, and
materials can cause damage to the device.
Daily
• Ensure exhaust hose is attached
Weekly
• Clean the exhaust outlet by removing the exhaust hose and wiping the space inside the exhaust outlet with a
paper towel or disposable cloth. To clean the exhaust hose, disconnect it from the blower exhaust duct, ush it
with water and allow to dry before reconnecting to the blower duct.
WARNING
Solvent can condense and collect inside the blower duct.
Annually
• Preventative Maintenance by CEM Certied Technician. Contact the CEM Service Department for assistance.
As Needed
• If installed, replace printer paper.