56
Maintenance Procedures
NOTE
QuikPrep temperature should always at 45°C when not in use for sample analysis.
• Fat Bias - This will set a bias for every result produced by the ORACLE if necessary to match historic data.
CEM does not endorse or recommend the use of a bias.
• Sample mass precision - This will adjust the precision of the sample mass gathered by the external
balance for calculating sample weight before ORACLE analysis.
• Normalization Data - This will produce the raw signal data from the Normalization tests performed during
setup. This should only be accessed at the request of a CEM trained specialist.
System
The “System” settings are used to record and control ORACLE information and connections, along with basic user
interface controls.
• Magnet S/N - This logs the serial number for the ORACLE magnet should for record keeping.
• Processor S/N - This is where the serial number for the ORACLE Processor box should be logged for record
keeping.
• System Time - This is used to set the correct time for the ORACLE so it is accurately reected in sample
logs.
• Volume - This adjusts the volume of the speakers for videos and other auditory cues on the ORACLE.
SMART 6 - This setting must be turned “ON” in order for the ORACLE to connect with a SMART 6 moisture/
solids analyzer. This setting should be “OFF” for high throughput users without a SMART 6.
• Automation: Use Robot - Setting this “ON” will allow the ORACLE to connect with the High Throughput
Robot option.
• Simulate Instruments - Turning this on will allow users to mimic the ORACLE performance without actually
running any samples, and is primarily used as a sales and demonstration tool.
• Service Mode - This will allow CEM trained specialists to have additional access to certain system
parameters and functions as necessary for servicing or optimizing the ORACLE.
• Screen Saver - This can be used to adjust the time in minutes until the ORACLE screensaver is activated
since the last time the touchscreen was touched.
• ‘Recent Results’ age lter - This will change the number of results that appear in the ‘Recent Results’
page.