114
32. Wait for the SMART System
5
to tare the
weight of the sample pads.
33. Lift the cover of the SMART System
5
.
Remove the square sample pads from the
balance pan.
34. Apply the sample in a thin, even layer to
one of the square sample pads and cover
the sample with the other square pad.
35. With the round pad still on the balance pan,
quickly and gently place the square sample
pads back on the balance pan. Close the
instrument cover.
36. Press READY.
Note: The T in the lower left corner of the
balance icon indicates that the weight of the
sample pads has been tared.
37. Press START.
The instrument reads and records the initial
weight of the sample prior to beginning the
analysis.
XXXXXX
Sample ID: xxxxxxxx
Power: xxx %
Temp (xxx): xx.x
Time: xx:xx
Wt: x.xxxxg
Press START to continue.
ID REDRY
1 2 3 4 5
T
READING BALANCE . . .
Modified Moisture/Fat
Spread sample on pads.
Place sample with pads on
balance and press READY.
ABORT
TARING BALANCE . . .

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CEM SMART System 5 and is the answer not in the manual?

CEM SMART System 5 Specifications

General IconGeneral
BrandCEM
ModelSMART System 5
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals