87
Note: The method test screen appears indicating
the power, temperature, and time.
15. If Autotrack is turned off and a method
identification name or number is required,
press the operation key below ID.
16. Using the operation keys below the arrows,
position the cursor on (select) the first
number or letter of the sample identifi-
cation. Press ENTER. Continue to position
the cursor on each letter or number and
press ENTER until the identification is
complete.
Note: If using an identification number only, use
the numeric keys to enter the number.
17. Press READY to return to the test screen.
18. Lift the cover of the SMART System
5
.
Place two glass fiber sample pads on the
balance pan. Close the instrument cover.
19. Press TARE.
20. Wait for the instrument to tare the weight
of the sample pads.
XXXXXX
Power: xxx %
Temp (xxx): xx.x
Time: xx:xx
Wt: x.xxxxg
Place pads on balance and
press TARE.
ID REWEIGH
1 2 3 4 5
SAMPLE ID:
0 1 2 3 4 5 6 7 8 9 A B
C D E F G H I J K L M N
O P Q R S T U V W X Y Z
Space Delete
Highlight choice, press ENTER.
Press READY to continue.
XXXXXX
Sample ID: xxx
Power: xxx %
Temp (xxx): xx.x
Time: xx:xx
Wt: x.xxxxg
Place pads on balance and
press TARE.
ID REWEIGH
1 2 3 4 5
TARING BALANCE . . .

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CEM SMART System 5 and is the answer not in the manual?

CEM SMART System 5 Specifications

General IconGeneral
BrandCEM
ModelSMART System 5
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals