EasyManua.ls Logo

claber Cometa - Test Mode; Systematic Valve Test Run

claber Cometa
15 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 11
For Cometa Evolution 4 Zone (p/n): 8014
For Cometa Evolution 6 Zone (p/n): 8016
TEST MODE
The TEST mode is particularly useful during installation or system maintenance as it allows all the valves to be
activated in sequence by the user; this way it is possible to check operation of the each part of the system.
- press the arrows a number of times together until TEST can be selected; the valve numbers are in a cycle.
- to activate the setting, press ENTER, appears OK
- press the arrows together to set the run time
- Press ENTER to conrm the setting; watering begins immediately.
- Press A and B at the same time to exit the function (by carrying out this operation, the watering test in
progress is NOT interrupted)
To stop watering after exiting the function ( A and B pressed at the same time), repeat the step 1 and then press
ENTER.
TEST mode takes priority: if it is activated, all other modes are interrupted (for example, MANUAL, AUTOMATIC,
SEMI-AUTOMATIC, etc.).
Cometa
Cometa
User manual
User manual
1
1
2
2
3
3
4
4
5
5
a
a
a
a
b
b
b
b
a
a
b
b
a
a
b
b
a
a
b
b

Related product manuals