Troubleshooting
Problem: Startup fails.
Check the external power supply and cables.
Check the power switch on the rear panel.
Verify that the SmartDrive is connected.
Problem: System will not power on.
Replace the external power supply.
Problem: Login fails.
Verify the user name.
Verify the PIN.
Verify that the user badge is correct and that its barcode quality is satisfactory.
Problem: The touch screen does not respond properly when touched.
Run the Calibrate Screen utility.
Problem: The formulary fails to load or is invalid.
A new formulary update package might have to be created and loaded. See your SLS
system administrator.
Problem: A drug container failed verification.
The drug might have to be added to or corrected in the formulary.
Make sure that the barcode on the drug is of good quality.
CAUTION: This is a serios issue. Notify your SLS system administrator.
Problem: A test label or syringe label did not print correctly.
Discard the label and try again.
If the label print quality is bad: Run the Clean Nozzles utility, Replace the ink cartridge, and
Replace the label media.
If the print is not aligned properly on the label, run the Adjust Label utility.
If the wrong drug information is printed on the label, the drug might have to be corrected
in the formulary. See your SLS system administrator.
CAUTION: This is a serious issue. Notify your SLS system administrator.
Problem: The barcode scanner is not scanning.
Make sure the barcode is correctly positioned. The red cross-hair should line up with the
barcode and the container or syringe should be as close to the front cover as possible.
Shutdown the system from the touch screen and then cycle power to the system.
Make sure the quality of the barcode is good.
Clean the scanner’s glass window.
The barcode symbology might not be supported. Contact Codonics Technical Support (+1
440.243.1198)
Problem: The label media is jammed.
Clear the label jam. Refer to “Clearing a Label Jam”.
Problem: The SLS will not connect to the network.
Verify that the Ethernet cable or Wi-Fi adapter is connected.
Verify that the SLS network settings are configured properly.
NOTE: For additional troubleshooting issues, refer to the Safe Label System User’s Manual.