61 Setting Delay Commands and Queries
Common Commands
DEL arg1 [ ps | ns ] Delay command set delay in active channel
DEL? Delay query both channels
ERR? or *ERR? Error query error status code
DEL arg1 Set Delay (QUERY)
Description
The del command sets the delay to the value specied in arg1. The desired delay must be within the total delay
range of the device.
Delay can be specied in either picosecond (ps) or nanosecond (ns) units if the corresponding ps or ns is
indicated. Default is picosecond units if not specied. The instrument will round DOWN to the nearest step
size resolution if unable to provide exact delay as entered.
Syntax
DEL arg1 [ ps | ns]
arg1 desired delay setting value
ps picosecond units
ns nanosecond units
Note: Picoseconds is the default units used if no units are specied in the command line.
Example
del 31250 ps
del 100
Returns None. Check error code.
DEL? Delay Setting? (QUERY)
Description Returns the current delay settings for Channel 1 and Channel 2
Syntax
<x.xxxxe-yy><nl>
(xxxx) is a 4 digit mantissa and (yy) a two digit exponent.
Returns 3.125000e-10
ERR? or *ERR? Error code query (QUERY)
Description Returns the error status
Returns
0 No Error
1 Invalid Command
2 Invalid Argument
3 No calibration
4 Delay setting limit (out of range)
5 Delay not set
19
COLBY INSTRUMENTS XT-100 VERSION 1.0 OPERATING AND PROGRAMMING MANUAL