MANUALP/4000_SERVICE_0712 July2012
26
DISPLAY PRESENT- BUT CONTROLLER WILL NOT OPERATE
CORRECTLY
- Data corrupted Switch off for ten seconds
Check set points and
calibration
- Component failure Call Service
DISPLAY ON BUT SHOWING RUBBISH!
The only power supplied to the display PCB is from the 5 volt supply on the
controller board. If the LED displays are illuminated it is a good indication that
the 5v power supply is present.
- Faulty Reset Turn Mains power off for 10
seconds
Call Service
CO
2
is too HIGH (IE Set for 5% but showing > 10% and not controlling).
- Check CO
2
Inlet Regulator Pressure is not set too high.
CO
2
Inlet Solenoid is Stuck OPEN. Call Service
ROUTINE MAINTENANCE
Suggested to be checked Annually:
Check glass door gasket for damage (cuts and splits), replace if necessary.
Check glass door seals correctly onto gasket, adjust side mounted door
latch plate if required.
Remove and clean water supply tray - refit.
Wipe out floor of cabinet to remove any water stains.
Check door controller overlay for damage, replace if necessary.
Check basic cabinet control functions and that user hi-limit control is set
appropriately.