Reviewing Day Statistics on the Display VFC400-USB User Guide 47
Data from 7 days ago
After pressing the button twice (skipping the
minimum statistics for day -6) the maximum
statistic from seven days ago is displayed:
l
The flashes to indicate that the data being
displayed is from seven days ago.
l
Seven days ago recordings above the upper
threshold were recorded, indicated by the
upper alarm threshold marker .
l
The duration was 5 hours 20 minutes.
l
An alarm event was generated, indicated by
the symbol. Neither duration from day -6 or
day -7 alone would cause an alarm event,
however the combined duration is 11 hours 50
minutes, which is more than the allowed 10
accumulative hours.
l
The maximum temperature recorded seven
days ago was 11.5°C.
Tip: Pressing and holding the button will get you back to the previous screen
for the same channel, rather than advancing one screen.