Performance Characteristics
Technical Specifications
Force FX Electrosurgical Generator C User’s Guide 9-7
Input Power
100–120 Volt 220–240 Volt
Maximum VA at nominal line voltage:
Idle: 52 VA
Bipolar: 450 VA
Cut: 924 VA
Coag: 530 VA
Maximum VA at nominal line voltage:
Idle: 52 VA
Bipolar: 450 VA
Cut: 924 VA
Coag: 530 VA
Input mains voltage, full regulation
range: 100–132 Vac
Input mains voltage, full regulation
range: 208–264 Vac
Input mains voltage, operating range:
85–132 Vac
Input mains voltage, operating range:
170–264 Vac
Mains current (maximum):
Idle: 0.4 A
Bipolar: 2.0 A
Cut: 7.0 A
Coag: 4.0 A
Mains current (maximum):
Idle: 0.2 A
Bipolar: 1.0 A
Cut: 3.5 A
Coag: 2.0 A
Mains line frequency range (nominal):
50 to 60 Hz
Mains line frequency range (nominal):
50 to 60 Hz
Fuses (2): F8 A Fuses (2): T4 A
Power cord: 3-prong hospital grade
connector
Power cord: 3-prong locally approved
connector