Standards and IEC Classifications
Technical Specifications
Force FX Electrosurgical Generator C User’s Guide 9-9
Standards and IEC Classifications
The Force FX-C generator meets all pertinent clauses of the IEC 60601-1 second edition
and IEC 60601-2-2 third edition.
Class I Equipment (IEC 60601-1)
Accessible conductive parts cannot become live in the event of a basic insulation
failure because of the way in which they are connected to the protective earth
conductor.
ATTENTION
Consult accompanying documents.
The generator output is floating (isolated) with respect to
ground.
Danger
Explosion risk if used with flammable anesthetics.
To reduce the risk of electric shock, do not remove the cover.
Refer servicing to qualified service personnel.
Non-Ionizing Radiation
Classified with respect to electrical shock, fire, and mechanical
hazards only in accordance with UL60601-1 and CAN/CSA
C22.2 No. 601.1.