EasyManua.ls Logo

Covidien Force FX-C - Output Characteristics; Maximum Output for Bipolar and Monopolar Modes

Covidien Force FX-C
130 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Output Characteristics
Technical Specifications
Force FX Electrosurgical Generator C User’s Guide 9-11
Output Characteristics
Maximum Output for Bipolar and Monopolar Modes
Power readouts agree with actual power into rated load to within 15% or 5 watts,
whichever is greater.
* An indication of a waveform’s ability to coagulate bleeders without a cutting effect
Mode Open Circuit
P–P Voltage
(max)
Rated Load
(max)
Power
(max)
Crest Factor*
Bipolar
Precise
Standard
Macro
450 V
320 V
750 V
100
Ω
100 Ω
100 Ω
70 W
70 W
70 W
1.5
1.5
1.5
Monopolar Cut
Low
Pure
Blend
1350 V
2300 V
3300 V
300
Ω
300 Ω
300 Ω
300 W
300 W
200 W
1.5
1.5
2.5
Monopolar
Coag
Desiccate
Fulgurate
LCF
Fulgurate
Spray
3500 V
8500 V
6900 V
9000 V
500
Ω
500 Ω
500 Ω
500 Ω
120 W
120 W
120 W
120 W
5
7.0
5.5
8

Table of Contents