Dakewe(Shenzhen)Medical Equipment Co,. Ltd 14 V4.1
3.2 Technical Parameters
Convention
Parameter
Nominal supply voltage
a.c.220V~240V/110V
Nominal frequency
50Hz/ 60Hz
Main fuse
4A
Nominal power
1200VA
Operating temperature range
+5°C to +30°C
Transport and storage temperature
+5°C to +55°C
Relative humidity for transport and storage
Maximum 60 %, non-condensing
IEC 61010-1 classification
Pollution level 2
Over-voltage category II
Atmospheric pressure
700 hPa ~ 1060hPa
A weighted noise level
≤ 70 dB (A)
Size and weight
Parameter
Size, without packaging (width × depth ×
height), unit: mm(Excluding hand wheel)
Size, without packaging (width × depth ×
height), unit: mm(Including hand wheel)
650 × 700 × 1280mm, ±5%
780×700×1280 mm, ±5%
Empty weight, without packaging
140kg
Weight with packaging
170kg
Section
Parameter
Section thickness range
0100μm
Section step amount
0μm20μm, with a step of 1μm
20μm50μm, with a step of 2μm
50μm100μm, with a step of 5μm
Section thickness deviation
±20%
Trimming thickness range
0600μm
Trimming step amount
1μm50μm, with a step of 5μm
50μm100μm, with a step of 10μm
100μm600μm, with a step of 50μm
Sample retraction
0~200μm adjustable, with a step of 5μm
Sample horizontal stroke
25mm±1mm
All relevant temperature indexes are only valid if the ambient temperature is
between 5°C and 30°C and the relative humidity does not exceed 60%.