Dakewe(Shenzhen)Medical Equipment Co,. Ltd 25 V4.1
5.1.4 Lighting and UV Lamp
Precautions:
1: UV disinfection can only be opened after the glass door is closed, and the UV lamp cannot
be operated when the glass door is open.
2: When the UV lamp is on, open the glass door and the UV lamp will automatically go out.
Close the glass door again and the UV lamp will automatically turn on. Every time you turn on
the UV lamp manually, the system defaults to turn on the UV lamp for 35 minutes, and it will
automatically turn off when the time is up.
5.1.5 Drying Button
1. Indication 1 is the Drying off state, only the function interface is displayed.
2. Indication 2 is the Drying open state, which is displayed on both the main interface and the
function interface.
Open the glass door and double-click the "drying" icon to turn on the drying function. When
the drying is turned on, it will run for 3.5 hours. After the time is up, the compressor and the
heating element will stop working, waiting for the operator to close the glass door or click
the "drying" icon to end the "drying" mode and enter the normal cooling state.
Precautions:
1: The "drying" function can only be turned on when the glass door is open. When the glass
door is closed when the "drying" function is running, the program will automatically end the drying
mode and return to normal cooling mode;
2: Before "drying" is turned on, all samples in the freezer must be emptied;
3: "Drying" is only suitable for a large amount of frosting and condensed water in the
equipment, and it is used when the water in the equipment needs to be eliminated as soon as
possible (it takes more than 24 hours to dry the water in the box after turning it off). It takes
about 6 hours to restart the “drying” function to the working temperature of -20°C.
During this period, the equipment cannot be used for frozen sectioning. Please select a
suitable time to enable the secondary function.