VII. Abnormalities and Troubleshooting
Dakewe (Shenzhen) Medical Equipment Co., Ltd. 79
www.dakewemedical.com
Cause analysis
The charging slot gap is larger in order to be compatible with most slides on the market.
Resolution
Log in to the root account, and enter the debug interface; increase the parameter of Right Ready Pos
by 0.3, and reduce the Left Ready Pos by 0.3.
12. Fail to start the instrument
Abnormality Description
No response after switching on the main power switch on the back of the instrument,and the instrument
does not start.
Cause analysis
Turn on the instrument within a short time after turning off the main power supply.
Resolution
Turn on the main power switch on the back of the instrument 1 min after turning off the main power
supply.