III. Initial Installation
10 Dakewe (Shenzhen) Medical Equipment Co., Ltd.
www.dakewemedical.com
(3) Unpacking and Installation
1. Unpacking requirements
The device needs to have enough space when being unpacked. The sides and the rear of the device
should be at least 1 m away from the wall.
To take the device out of the pallet, always reserve at least 1.5 m space in front of the device, and
move the device packaging box as close to the final installation site as possible.
At least two people are required to unpack and carry down the device.
If there is a large temperature difference between the warehouse and the
installation site and the air humidity is high, condensation may form inside the
device. In this case, you must wait for at least two hours before turning on the
device, otherwise the device would be damaged.
Before unpacking, please check that the packaging box is intact, and no damages
or deformities in appearance. In case of any of the conditions above, please mark on
the shipping documents accordingly.
To ensure the device will operate properly, the device should be installed with at
least 100 mm space around the device for ventilation.
Packaging
box
1.5 m
1 m
1 m
Figure3- 5Diagram of Unpacking Area
Figure3- 6Removal Tool
2. Remove the wooden crate
The following steps are required for removing the crate:
(1) Remove the fasteners using a disassembly tool or a flat-blade screwdriver (Figure 3-6).
(2) Align the disassembly tool (or flat-blade screwdriver) with the round hole at the bend of the
fastener and press in the direction of the force point (Figure3-8).
(3) Remove the foam in the box, and take the instrument out.
(4) To reassemble the crate: press the fasteners into the slots on the box based on the length direction.
Figure3- 7 Fasteners
Figure3- 8Fastener Press-fit Point
1 m