Appendix 5: Electromagnetic Compatibility Indicators
Dakewe (Shenzhen) Medical Equipment Co., Ltd. 85
www.dakewemedical.com
Appendix 5: Electromagnetic Compatibility Indicators
Note:
[Slide Printer] shall meet the emission and immunity requirements specified in
IEC61326-2-6, as shown in the table below.
The user shall be responsible for ensuring the electromagnetic compatibility of the environment for the
equipment to work normally.
It is recommended to evaluate the electromagnetic environment before the equipment is used.
Warning:
[Slide Printer] is designed and tested according to Class A equipment in CISPR 11. In the home
environment, the equipment may cause radio interference, so that protective measures should be taken.
It is prohibited to use the equipment next to an intense radiation source (such as an unshielded RF source),
otherwise the Ordinary operation of the equipment will be affected.
Table I: Electromagnetic emission
CISPR 11 conducted emission
CISPR 11 radiated emission
IEC61000-3-2 harmonic emission
IEC61000-3-3 voltage fluctuation/scintillation
emission
Table II: Electromagnetic immunity
Electrostatic discharge
(ESD)
Contact discharge: ± 4 kV
Air discharge: ±2 kV, ± 4 kV, ± 8 kV
Radio frequency
electromagnetic field
80MHz-6.0GHz,3V/m
AM80% 1kHz, step 1%;
Dwell time 1s
Power cord: ±1 kV (5/50 ns, 5 kHz)
Line-to-Earth: ± 1 kV
Line-to-Neutral: ± 0.5 kV
Radio frequency
transmission
Power switch: 3 V, 150 kHz- 80 MHz,
80 %AM
Power frequency magnetic
field
Voltage sag and interruption
0.5 cycle 0%;
1 cycles 0%;
25/30 cycles 70%;
250/300 cycles 0%
Performance criteria:
A. In the test, the performance is Ordinary within the specified limit.
B. In the test, the function or performance is degraded or lost temporarily, but can restore automatically.
C. In the test, the function or performance is degraded or lost temporarily, and cannot restore until reference or
system operation by the operator