Avance
5-6 M1204616
Machine check
The Machine Check or the individual checks must be performed at
least once within every 24-hour period.
Perform the Machine Check at the start of each day. The machine
check runs automatically and beeps to indicate when it is finished or if
interaction is required.
The Machine Check does a Machine Check-System check,
Machine Check-Circuit check, and a Machine Check-Circuit O2
cell check (if circuit O2 cell is present). When one of the checks is
completed, the next check begins.
1. Turn the System switch to On.
2. Select Machine Check and follow the instructions.
3. If a check fails, follow the instructions to perform a recheck or
accept the results.
4. When the Machine Check is completed, start a case.
Note In case of a patient emergency, the Machine Check may be
bypassed by selecting Start Case from the Checkout menu. The
general message ‘Please Do Checkout’ is displayed if a Machine
Check is not completed with passing results within 24 hours.
Machine check -
system
The Machine Check-System checks the Bag/Vent switch, proper
gas supply pressures, ventilator operation and leak, battery and
electrical power, circuit compliance, and flow control operation. This
is a two-step check.
1. Set the Bag/Vent switch to Vent.
2. Open the patient Y.
3. (ACGO option only.) Set the ACGO switch to Circle.
4. Calibrate the flow sensors. Make sure the flow sensors are
securely latched when they are reinserted.
5. Select Start. The display shows the checks being run.
The system beeps when this portion of the check is done.
The results are shown on the display.
6. Make sure the bellows is fully collapsed.
7. Occlude the patient Y.
8. Select Continue. The display shows the checks being run.
9. When the check passes, the next check starts.
Machine check -
circuit
The Machine Check-Circuit checks the Bag/Vent switch, proper gas
supply pressures, airway pressure measurement transducer, APL
valve, and manual circuit leak.
1. Occlude the patient Y.
2. Set Bag/Vent switch to Bag.
3. Set the APL valve halfway between 30 and 70.