4 Repair
4.3 Replacing parts (without removing covers)
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
94
66 81 832 D3696
D3696.076.01.02.0205.2019
4.3.2.4
Replacing the PC including the power supply unit
Removal
1.
Open the service cover (see "Opening the service cover  [→52]").
2.
If the option is available, remove the battery (see "Replace battery
(optional)  [→100]").
3.
Unscrew the partition plate (2 screws).
4.
Remove the partition plate (as shown).