0044-0391-0M Rev A y solo Digital Clinical Scale
19
CALIBRATION, CONT.
!" (Calibrate Scale)
Step 1. The display will change to show !".
Step 2. To skip calibration and save the previous settings, press the
key to exit
calibration. The scale will turn off automatically. Proceed to Step 14.
Step 3. Otherwise, press the ©
©
or ª keys to select ( (Y = start calibration). Note that
you can still select 0 (n = exit calibration), skip calibration and turn the scale off.
Step 4. Press the
key.
Step 5. The display will change to show the internal count.
Step 6. Make certain the scale platform is empty and free of debris.
Step 7. Press the
key to save the zero reference point.
Step 8. The display will change to show the first calibration point  kg or  lb
Step 9. Place the first calibrated test weight on the scale platform.
Step 10. When reading is stable, the first weight reading will be saved and the display will
change to show the second calibration point  kg or  lb.
Step 11. Place the next calibrated test weight on the scale platform.
Step 12. When reading is stable, the second weight reading will be saved.
Step 13. The display will change to show $** and then automatically shut the scale off.
Step 14. The scale is now ready for operation.