0044-0391-0M Rev A y solo Digital Clinical Scale
CALIBRATION
Your solo Digital Clinical Scale was calibrated at the factory and should not require
adjustment. In the event that the scale should need re-calibration, the following describes
the calibration procedure. To maintain the instrument’s high degree of accuracy, a qualified
technician should perform this function.
Before beginning calibration, the following equipment is required:
150 kg or 300 lb of calibrated test weight
To Enter Calibration
1. With the scale off, press and hold the “ ” and “ ” keys and then press the
“
” key to turn on the scale.
2. The display will turn on and show *;.
3. Release all keys.
4. The scale is now ready for calibration.
*; (Set Calibrated Gravity Constant)
This setting (e.g. )'(') is the acceleration of gravity constant value for the location where
the scale was calibrated.
With the display showing *&, press the “
” key to show current setting. If the value
displayed is acceptable, press the “
” key and proceed to the next prompt. Otherwise,
press the ©
or ª keys to adjust the setting then press the “ ” key to save it and
proceed to the next prompt.
Allowable values are 9.700 through 9.900.
! (SET OPERATED GRAVITY CONSTANT)
This setting (e.g. )) is the acceleration of gravity constant value for the location where
the scale will be operated.
With the display showing !, press the “
” key to show current setting. If the value
displayed is acceptable, press the “
” key and proceed to the next prompt. Otherwise,
press the
or ª keys to adjust the setting then press the “ ” key to save it and
proceed to the next prompt.
Allowable values are 9.700 through 9.900.
IMPORTANT! When a prompt and value displayed are acceptable, press
the “ ” key to save the setting and proceed to the next prompt. To
change a setting, press the
or ª keys to adjust the setting (or select a
new value) and then press the “
” key to save the new setting and
advance to the next prompt.