0044-0391-0M Rev A y solo Digital Clinical Scale
SERIAL DATA FORMAT
The serial port on the solo Digital Clinical Scale is a bi-directional port and can receive and
transmit data. It can be used to communicate to a computer by to pressing the “ ” key
or transmitting a command to the scale.
NOTE: The serial output parameters are fixed at 9600 baud, with the data format at
8 bits, No parity, and 1 stop bit.
To connect to a computer, you will need a 9-Pin serial interface cable (Detecto part number
6600-1891) which is connected to the serial port located under the small cover on the back
of the display.
Output Format
The format below represents the output directly to a computer for processing of data and
includes commas between each field and a carriage return to terminate the record.
Time, Date, BMI, Height, Weight <CR>
Remote Print Command
The host device (computer) sends:
<LF> P <CR>
This command is equivalent to pressing the “ ” key. The scale will respond by sending
the above Output Format to the computer.
12:00,08/07/15,21.5,5’03.0”,132.3lb