24
A-DV56
EN
troublesHooting
Mask and Headgear
Clean the mask and headgear according to the manufacturer’s instructions.
TRoUblesHooTinG
DAnger
Electric Shock Hazard - Do not remove the device cabinet; there are no
user-serviceable internal components. The cabinet should only be
removed by a qualified DeVilbiss provider.
If you are having trouble with the operation of your DeVilbiss IntelliPAP, please refer to the
table below for a list of problems with possible causes and remedies. If you are unable to
resolve the problem by following the instructions, contact your home care provider for
assistance.
Problem Probable Cause Remedy
Nothing shows
on the display.
1. Device is not plugged in or
the power cord is not fully
inserted.
2. There is no outlet power.
1a. Verify that power cord is firmly
connected to the IntelliPAP and the
power outlet.
1b. If you are using a DC power source
make sure the cable connections
are secure. Verify that your battery
is charged.
2. Locate a suitable functioning power
source.
The IntelliPAP
does not start
when breathing
into the mask.
1. The Auto-ON feature has
been disabled.
2. There is no power to the
device.
3. Breath is not deep enough
for the Auto-ON feature to
detect.
4. You are using a full face mask
that has an anti-asphyxia
valve.
5. The air supply port plug is
missing or not fully inserted.
6. The humidifier chamber is
not fully engaged into the
cradle or is missing.
1. Use the ON/OFF button to start and
stop the device.
2. Verify that the power cord is firmly
connected to the device and the
power outlet.
3. Breathe deeply in and out to start
the IntelliPAP
4. Auto-ON may not work because
your exhaled breath escapes through
the open valve. Use the ON/OFF
button to start and stop the device.
5. Make sure the air supply port plug is
fully inserted into the unit.
6. Slide the humidifier chamber fully
into the slot. Verify the latching lever
snaps into place.