26
A-DV56
EN
specificAtions
Problem Probable Cause Remedy
Air from ow
generator seems
to be too warm.
1. The air filter(s) is/are dirty.
2. The air inlet port is blocked.
3. The room temperature is too
high.
4. The IntelliPAP is located near
a heater.
1. Refer to the Cleaning section for
instructions on filter maintenance.
2. Unblock the air inlet.
3. Reduce the room temperature.
4. Locate the device away from the
heat source so the air being drawn
into the device is at room
temperature.
If the problem persists contact your
home care provider.
Nasal, sinus or
ear pain,
possibly with
runny nose.
You may have a reaction to air
ow pressure.
Discontinue use and contact your
physician.
Bilevel changes
pressure too
early or too late
(not following
breathing)
1. Trigger Sensitivity is not set
correctly
2. Backup rate (time) is too
high, the patient is breathing
slower than the backup rate
1. Adjust the inspiratory and expiratory
trigger sensitivity to adjust triggering
for the patient
2. Reduce Backup Rate setting
Bilevel changes
pressure too
early or too late
(not following
breathing)
1. Trigger Sensitivity is not set
correctly
2. Backup rate (time) is too
high, the patient is breathing
slower than the backup rate
1. Adjust the inspiratory and expiratory
trigger sensitivity to adjust triggering
for the patient
2. Reduce Backup Rate setting
sPeCifiCaTions
Size ........................ 4.2” H x 6.5” W x 6.9” D (10.7 cm x 16.5 cm x 17.5 cm)
Weight ......................................................2.7 lbs. (1.22 kg)
Electrical Requirements AC ................................. 100-240V~, 50/60 Hz
Electrical Requirements DC................................ 10.4 - 15 VDC, 5 Amps
Maximum Power Consumption ..................65 watts max from AC power source
(flow generator only)
Bilevel Pressure Range .............................................3-25 cmH2O
Operating Temperature Range...........................41°F to 104°F (5°C to 40°C)
Operating Humidity Range.......................... 0% to 95% RH non-condensing
Operating Atmospheric Conditions ....................Sea level to 8,500 feet (2600 m)
Storage & Transportation Temperature Range...........-40°F to 158°F (-40°C to +70°C)