Carrying out the cleaning
26
Nebuliser chamber
1
(brownish colour)
In warm
water
with mild
household
cleaner
Immerse
in diluted
solution
Rinse
cycle
up to
95° C
Up to
134° C
max�
18 min
Nebuliser cover
1
(white)
Nebuliser chamber
1
with quartz module
Tubing Not per-
mitted
Nebuliser chamber
1
(transparent)
In warm water
with mild house-
hold cleaner
Immerse
in diluted
solution
Rinse
cycle
up to
95° C
Up to
121° C
max.
20 minNebuliser chamber
cover
1
(transparent)
Bacteria and
virus filter
Single use items: Replace after 1 week at the latest or in
accordance with the manufacturer’s instructions� When using a
closed sterile water system, it is recommended that you replace
the filter after 48 hours of operation or every 14 days�
Medicine cup Only to be used once
Medicine cup cover
Housing Wipe with a
damp cloth
Wipe over
disinfectant
Not per-
mitted
Not per-
mitted
Power cord
Accessory kits
1� After each use, rinse the nebuliser chamber and nebuliser chamber cover with distilled
water� Carry out the disinfecting process with standard medical disinfectants�
The quartz module remains screwed into the nebuliser chamber.
2� After the recommended period of time (in accordance with the manufacturer’s
instructions), use a damp cloth to clean off any remaining disinfectant from the
components and then dry them�
3� Caution: heated steam accelerates the natural ageing process of plastics�
Changes in material can affect the functionality of the device parts�
4� Remove the deposits or solids from the float chamber, valve seat and valve unit with
tweezers� Rinse all parts thoroughly under running water�
Components Cleaning Disinfecting
2
Dish-
washer
Hot steam
sterilisation
3
English