Technical data
Ambient conditions
During operation: ���������������������������� Temperature 10° to 40° C, Humidity 0 to 90 %
relative humidity without condensation
During storage: ������������������������������ Temperature -20° to 70° C,Humidity 0 to 90 %
relative humidity without condensation
Performance data
Voltage ��������������������������������������������� 110 V / 127 V / 230 V; AC, 50 to 60 Hz
Fuses � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � IEC 127-T800 L 250 V, 2 EA
Power consumption: ����������������������� Ultrasonic nebulizer 50 VA; Hose heating 30 VA
Resistance of hose heating ��������������������������������������������������� 3 Ohm (± 5 %)
Ultrasonic frequency ����������������������������������������������������� 1,68 MHz (± 5 %)
Noise pressure level ������������������������������������������������������� about 35 dB(A)
Humidity protection type �������������������������������������������� IP 32 (dripping water)
Method of operation: Continuous operation
Nebulizer output ���������������������������������������������������������� max� 3 ml/min
Nebulizer output ������������������������������� with sterile water system up to 2,5 ml/min
Droplet size (MMAD) ��������������������������� average 4 µm, about 86 % lower than 5 µm
Air flow ������������������������������ up to 20 L/min (when using AQUA+ F or SecuRed Big)
Classification
as per EC Directive 93/42/EEC, Annex IX Class IIa
CE-Labelling
Directive 93/42/EEC concerning Medical Devices
7. Technical data
31
Medical electrical equipment needs special precautions regarding electromagnetic compa-
tibility (EMC) and needs to be installed and put into service according to the EMC information
provided in the technical documentation avail-able from DeVilbiss Healthcare Service upon
request�
Portable and mobile RF communications equipment can affect medical electrical equip-
ment.
Pins of connectors identified with the ESD warning symbol shall not be touched and not be
connected unless ESD precautionary procedures are used� Such precautionary procedures
may include antistatic clothing and shoes, the touch of a ground stud before and during
connecting the pins or the use of electrically isolating and antistatic gloves� All staff involved
in the above shall receive instruction in these procedures�
Dimensions:
Height over all �������������������������������������������������������������������� 19,0 cm
Width ���������������������������������������������������������������������������� 20,5 cm
Depht ���������������������������������������������������������������������������� 31,5 cm
Weight ����������������������������������������������������������������������� about 3,5 kg
General information on electromagnetic compatibility (EMC) according to
the international EMC standard IEC 60601-1-2: 2001
English