Operator’s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc.
- 20
blocked.
Alarm not cleared
Press“ ”to clear the alarm.
“ ”
Pump door is open. Close and lock the pump door
IV Tubing folded Check whether the IV tubing is folded
Roller clamp is
closed
Open the roller clamp
needle is blocked Check whether the needle is blocked
“ ”
Occlusion in
IV tubing
The IV set is too
hard or large
Replace the IV set or lower down the sensitivity
of occlusion
1. Battery voltage is low
When running on battery, indicates
that battery voltage is low. Promptly connect to AC
power or replace with same type battery .
“ ”
2.Battery failed
When running on AC, indicates that
battery failed. Check the connection of the
battery or replace with same type battery.
Infusion finished
1.Change to new solution container,
and set the parameter again . 2.Infusion is
finished, remove the IV set.
“ ”
Auto-adjustment finished
Turn off the machine and restart .(The
“ ” flashes)
Motor rotation abnormal Please stop to use and contact customer service