11.1.3 Notice
The centrifuge rotor can separate sample which density lower than 2.0g/ml. if the separated samples
density is over 2.0g/ml, please calculate allowable speed depending on the following formula.
Allow Speed (rpm)= Maximum speed×(2.0(g/ml)/Sample density (g/ml))
1/2
11.1.4 Autoclaving
A12-2 rotor is manufactured in high-strength aluminum alloy material and can be autoclaved: 121℃ (1.0kg/cm2),
20 minutes.
A12-2P rotor is made of plastic, cannot be high-pressure sterilization and UV irradiation, only ordinary
sterilization can be used.
CAUTION:
The lid of the rotor is made of plastics, can not be high-pressure sterilization, only ordinary sterilization
can be used.
11.2 Tubes
11.2.1 Cleaning and sterilizing tubes
Table 11.2 Cleaning and sterilizing conditions for tubes
O:Applicable X:Inapplicable