20 Supplement | Area Control Unit
Reprocessing
7.3.2 Classification of device-specific components
The following classification is a recommendation from Dräger.
Non-critical
–Surface
7.4 Validated reprocessing procedures
7.4.1 Overview of the reprocessing procedures of the components
7.4.2 Surface disinfection with cleaning
Components:
–Surfaces
Prerequisites:
The surface disinfectant has been prepared in accordance with the
manufacturer's instructions.
The manufacturer's instructions, e.g., regarding shelf life or application
conditions, are observed.
An uncontaminated, lint-free cloth soaked in surface disinfectant is used for the
cleaning surface disinfection.
Cleaning
1. Wipe off obvious soiling with a disposable cloth soaked in surface disinfectant.
Dispose of the cloth.
2. Wipe all surfaces. After that, there must no longer be any soiling visible.
Surface disinfection
3. Wipe cleaned surfaces again to visibly wet all surfaces to be disinfected with
surface disinfectant.
4. Wait for the surface disinfectant contact time.
5. At the end of the contact time, moisten a new, uncontaminated and lint-free cloth
with water (at least drinking water quality).
6. Wipe all surfaces until no remains of the surface disinfectant, such as foam
residues or streaks, are visible.
7. Wait until the surfaces are dry.
8. Check the surfaces for visible damage and, if necessary, replace the product.
Components Surface disinfection
with cleaning
Description of the procedure
Surface Yes (see "Surface disinfection with
cleaning", page 20)
Surface disinfec-
tant
Manufacturer Concentration Contact time
Dismozon plus BODE Chemie 1.6 % 15 min
Oxycide Ecolab USA 2.3 % 5 min