Supplement | Area Control Unit 11
Overview
4 Overview
4.1 Area control unit with digital display
By default, the area control unit is delivered with a display unit (pressure gauge and
LEDs). The display unit is supplemented by a digital display. The digital display is
integrated into the door and equipped with a door opening alarm.
The following information is displayed on the backlit color touchscreen of the digital
display.
– The numerical values of the current pressures of all monitored gases
– The visualization of pressure alarms
– The numerical values of the current volume flows of all monitored gases
Display options for the monitored gases
– AIR, additional supply circuits if required (AIR (1), AIR (2), AIR (3))
– AIR800
– O
2 additional supply circuits if required (O2 (1), O2 (2), O2 (3))
– CO
2
– N2O
– N
2O+O2 (mixture of nitrous oxide and oxygen)
– VAC