24 Supplement | Area Control Unit
Technical data
9 Technical data
9.1 Ambient conditions
During operation
Temperature 10 to 40 °C
(50 to 104 °F)
Air pressure 800 to 1060 hPa
(1950 - 0 m NN)
Relative humidity 30 to 95 % (non-condensing)
Pollution level of the environment 2
Maximum volume < 80 db
During storage and transport
Temperature -20 to 60 °C
(-4 to 140 °F)
Air pressure 800 to 1060 hPa
Relative humidity 30 to 95 % (non-condensing)
9.2 Electronic operating specifications
Mains power supply
Electrical mains connection 230 V
50 Hz
Permissible voltage fluctuation of the mains
power supply
± 10 %
Overvoltage category II
Current consumption
At 230 V Max. 0.2 A
Max. 0.33 A (with AIM power supply unit)
Power consumption
With AIM Max. 76 VA (at 230 V)
Without AIM Max. 46 VA (at 230 V)
Protection class Class 1
Door opening alarm
Normally open or normally closed Open in the event of an alarm
Switching voltage Max. 50 V DC
Switching current Max. 1 A