Fabius
®
Repair Instructions
3-83
Copyright 2000 Draeger Medical, Inc.
sec3_repairinstruc_c.fm
15.3 Removing the Gas Box
• Switch control box off using ON/OFF switch (ON/OFF switch is located on rear
panel of control box).
• Disconnect power plug from AC outlet.
• Disconnect gas supply hoses from pipeline system.
• Remove fixing screws 1 from cover 2 and pull cover 2 slightly out.
Fig. 68: Rear view of Fabius, removing gas box
1
1
1
2
RETURN TO CD-ROM TABLE OF CONTENTS