Fabius
®
Repair Instructions
3-23
Copyright 2000 Draeger Medical, Inc.
sec3_repairinstruc_c.fm
13 DrägerService Mode (software 2.00)
DrägerService Mode allows the following actions:
calibrating the pressure sensor (Paw)
calibrating the PEEP valve
deleting error log entries
reading out/resetting elapsed hours meter
selecting the function (monitor or ventilator or monitor and ventilator)
initializing Fabius (calibration values, elapsed hours, etc.)
13.1 Accessing the DrägerService Mode (software 2.00)
The DrägerService Mode can only be accessed while Fabius is in the standby mode.
Press the Menu key.
Select keycode by turning rotary knob.
Confirm by pressing rotary knob.
The respective number is displayed on black background.
Select keycode 112 by turning rotary knob clockwise.
Confirm by pressing rotary knob.
An additional screen page is displayed.
13.2 Exiting the DrägerService Mode (software 2.00)
Switch the control box off.
The DrägerService Mode is used to calibrate critical
components (Paw sensor, PEEP valve). The DrägerService
Mode should only be used by qualified and authorized
DrägerService technicians.
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS