6 Debug Mode (software 1.00)
The following information is displayed in the debug mode during normal operation of
Fabius: system data, measured values, error log. One debug mode page is available.
System data cannot be changed.
6.1 Accessing the Debug Mode (software 1.00)
The debug mode can be accessed from any operating mode.
Press the Menu key.
Select keycode by turning rotary knob.
Confirm selection by pressing rotary knob.
The selected numeric code is displayed on black background.
Select keycode 1 by turning rotary knob one position clockwise.
Confirm selection by pressing rotary knob.
Press rotary knob again.
An additional screen page is displayed. Fabius remains in the normal mode and can
be operated as desired.
6.2 Exiting the Debug Mode (software 1.00)
Press the Menu key.
Select keycode by turning rotary knob.
Confirm selection by pressing rotary knob.
The selected numeric code is displayed on black background.
Select keycode 0 by turning rotary knob one position counter-clockwise.
Confirm selection by pressing rotary knob.
Press rotary knob again.
Fabius is now in the normal operating mode.