Fabius
®
Repair Instructions
3-18
Copyright 2000 Draeger Medical, Inc.
sec3_repairinstruc_c.fm
11 Error Log
Error Error Description Possible Cause
0 No error
11 Reset; cause unknown Hardware error. Electromagnetic
interference. EEPROM needs
replaced
14 RAM (Control PCB) RAM needs replaced.
15 EPROM (Control PCB) EPROM needs replaced.
16 Watchdog reset (Control PCB) Hardware error. Software error.
Watchdog time exceeded.
17 A/D converter Error during A/D conversion.
18 Graphics display Graphics display or connector
needs replaced.
21 Battery Wrong battery installed or
installed battery flat.
22 Analog voltage Error in ±12 V voltage supply.
Control PCB needs replaced.
23 Supply voltage Battery exhausted during power
supply from battery.
31 EEPROM* Read/write error during system
start-up. EEPROM needs
replaced. Data error.
32 EEPROM* write error Error during EEPROM writing.
EEPROM needs replaced. Data
error.
33 EEPROM* read error Error during EEPROM reading.
EEPROM needs replaced. Data
error.
34 RAM backup error Error while reading internal RAM
data during system start-up.
35 Data check error/system reset
occurs
Checksum of settings or
operating mode incorrect.
36 Communication error RAM locations inverted or needs
replaced.
37 Calibration value check error (O
2
sensor, pressure sensor, flow
sensor, PEEP valve)
EEPROM error. RAM error.
Standard values are being used.
Calibration required.
* Later models do not have an EEPROM and require board replacement.
Rev. A
RETURN TO CD-ROM TABLE OF CONTENTS