6.3 Debug Mode Table (software 1.00)
Table 3: Debug Mode (software 1.00)
Pressure sensor value (Paw) in mbar
(default)
P XXX (0 to 738)
Plateau pressure in mbar
Plat XXX (varies with system
settings)
Flow sensor A/D converter value (default)
FL XXX (0 L/min 512)
Supply voltage (approx. 27 V) in volts
U_24 XX.X (22.5 35.0)
Oxygen sensor channel 1 in mV
O2_1 XXX (9 20)
Supply voltage (+12 V) in volts
U+12 XX.X (12.0 ±2.0)
Oxygen sensor channel 2 in mV
O2_2 XXX (9 20)
Supply voltage (12 V) in volts
U12 XX.X (12.0 ± 2.0)
Expiratory Vt during inspiration (mL)
Vtix XXX (varies with system settings,
ideally 0)
EEPROM error code
EEPR XXXX
Pressure sensor calibration value
CalP XXX (704 768)
Battery charging current in amperes
LADI X.X (approx. 0.0 2.0)
Flow sensor calibration value
CalF XXX (482 542)
Motor current in amperes
MOTI X.X (0.0 2.5)
Oxygen sensor calibration value (21% O2)
CalO XX (39 113)
Pressure between rolling seals in mbar
PMEM XXX (200 ±50)
Monitoring error code
MONI XXXX
Ventilation error
VENT XXXX
Calculation from controller data (set values):
Tidal volume (mL) frequency (1/min)
V: XXXX F: XX
System temperature in °C
T XX (max. 60°C)
Error log
(most recent error entry on extreme left)
LOG
XX XX XX XX XX XX XX XX XX
RETURN TO CD-ROM TABLE OF CONTENTS