Fabius
®
Repair Instructions
3-9
Copyright 2000 Draeger Medical, Inc.
sec3_repairinstruc_c.fm
6.3 Debug Mode Table (software 1.00)
Table 3: Debug Mode (software 1.00)
Pressure sensor value (Paw) in mbar
(default)
P XXX (0 to 738)
Plateau pressure in mbar
Plat XXX (varies with system
settings)
Flow sensor A/D converter value (default)
FL XXX (0 L/min 512)
Supply voltage (approx. 27 V) in volts
U_24 XX.X (22.5 – 35.0)
Oxygen sensor channel 1 in mV
O2_1 XXX (9 – 20)
Supply voltage (+12 V) in volts
U+12 XX.X (12.0 ±2.0)
Oxygen sensor channel 2 in mV
O2_2 XXX (9 – 20)
Supply voltage (–12 V) in volts
U–12 XX.X (12.0 ± 2.0)
Expiratory Vt during inspiration (mL)
Vtix XXX (varies with system settings,
ideally 0)
EEPROM error code
EEPR XXXX
Pressure sensor calibration value
CalP XXX (704 – 768)
Battery charging current in amperes
LADI X.X (approx. 0.0 – 2.0)
Flow sensor calibration value
CalF XXX (482 – 542)
Motor current in amperes
MOTI X.X (0.0 – 2.5)
Oxygen sensor calibration value (21% O2)
CalO XX (39 – 113)
Pressure between rolling seals in mbar
PMEM XXX (200 ±50)
Monitoring error code
MONI XXXX
Ventilation error
VENT XXXX
Calculation from controller data (set values):
Tidal volume (mL) frequency (1/min)
V: XXXX F: XX
System temperature in °C
T XX (max. 60°C)
Error log
(most recent error entry on extreme left)
LOG
XX XX XX XX XX XX XX XX XX