Fabius
®
Function Description
2-10
Copyright 2000 Draeger Medical, Inc.
sec2_funcdescrip_b.fm
If the O
2
-flush button is pressed, the fresh-gas ejector prevents the fresh gas from
flowing back into the anesthetic vaporizer. This avoids an increase in the anesthetic
gas concentration.
Fig. 9: Function diagram of the gas box (3-gas version), part 2
O
2
flush
anesthetic
vaporizer
fresh gas
fresh-gas