15. Leak test
Perform leak test between gas source and
flow control valve.
Test prerequisites:
All gas supplies depressurised.
Flow control valves closed.
15.1 O2-air/N2O-CS operation
Connect CS connection hoses with test
adapter 7901495 to test cylinder (test
pressure reducer).
Test set-up:
Build up test pressure of 5 bar.
All flow control valves at anaesthesia
machine closed.
Close cylinder valve following pressure
equalisation.
Test values:
Cylinder pressure indicated (high-
pressure gauge of test cylinder) may not
drop by more than 10 bar within 2 minutes.
☞