Dräger Interlock
®
5000 / 7000
i
en
Instructions for Use
3
de
Gebrauchsanweisung
20
sv
Bruksanvisning
37
fr
Notice d’utilisation
54
fi
Käyttöohje
71
no
Bruksanvisning
88

Table of Contents

Other manuals for Dräger Interlock 5000

Questions and Answers

  • A
    Alex AdamsSep 3, 2025
    How to fix 'Blow harder' on Dräger Safety Equipment?
    • K
      Katherine WilliamsSep 3, 2025
      If the Dräger Safety Equipment displays 'Blow harder!', it means the breath sample was blown too softly. Repeat the test, blowing harder this time.
  • M
    Michael LoweSep 1, 2025
    What does 'Please retry' mean on Dräger Safety Equipment?
    • E
      enealSep 2, 2025
      If the Dräger Safety Equipment displays 'Please retry', it means a fault was detected during the delivery of the breath sample. Deliver the breath sample again when requested by the instrument.
  • A
    Annette DeckerAug 31, 2025
    What to do if Dräger Safety Equipment displays 'High Temperature'?
    • S
      Shaun JamesAug 31, 2025
      If the Dräger Safety Equipment displays 'High Temperature', it means the handset temperature is too high. Detach the handset and take it to a place with a moderate temperature. Connect the handset again and perform the test.
  • C
    Curtis WongAug 29, 2025
    How to fix 'Low Temperature' error on Dräger Safety Equipment?
    • J
      jessica48Aug 29, 2025
      If the Dräger Safety Equipment displays 'Low Temperature', it means the handset temperature is too low. Detach the handset and take it to a place with a moderate temperature. Connect the handset again and perform the test.
  • J
    James HernandezAug 28, 2025
    What to do if Dräger Safety Equipment shows 'Control unit error'?
    • V
      Victoria RussellAug 28, 2025
      If the Dräger Safety Equipment displays 'Control unit error', it indicates an error in the control unit. Try switching the ignition off and on, then repeat the test. If the error message reappears, further assistance is needed.
  • J
    Joy RodriguezAug 27, 2025
    Why is there no display message on my Dräger Safety Equipment handset?
    • J
      Joshua PrattAug 27, 2025
      If there is no display message on the Dräger Safety Equipment handset, it indicates no connection between the handset and control unit. Check the connecting cable plugs, disconnect the plug from the handset, and reconnect it. If no messages are still shown, further assistance is needed.
  • K
    Kristin MyersAug 25, 2025
    What does 'Handset error' mean on Dräger Safety Equipment?
    • J
      james35Aug 25, 2025
      If the Dräger Safety Equipment displays 'Handset error', it indicates a fault on the handset. Try switching the ignition off and on, then repeat the test. If the error message reappears, further assistance is needed.
  • J
    Jason RodriguezAug 22, 2025
    How to fix Dräger Safety Equipment when it says 'Check cable and reconnect'?
    • H
      Hannah MccormickAug 23, 2025
      If the Dräger Safety Equipment displays 'Check cable and reconnect!', it means there is no connection between the handset and control unit. Check the plugs of the connecting cable and reconnect them to the handset and control unit. If the error message reappears, further assistance is needed.

Summary

1 For Your Safety

1.1 General Safety Statements

General safety statements and precautions for using the Dräger Interlock product.

2 Product Description

2.2 Functional Description

Detailed explanation of how the Dräger Interlock 5000/7000 operates and its features.

2.3 Intended Use

Defines the primary purpose of the Dräger Interlock 5000/7000 as a breath alcohol measuring instrument.

3 Using the Dräger Interlock

3.1 Installation

Instructions and warnings related to the installation of the Dräger Interlock system.

3.3 Making a Measurement

Step-by-step guide on how to perform a breath alcohol measurement.

4 Menu Options

5 Troubleshooting

Display Issues and Causes

Troubleshooting common display errors and their potential causes.

Breath Sample Issues

Resolving issues related to breath sample quality and delivery during measurements.

Instrument Status Errors

Addressing errors like 'Test failed', low/high temperatures, or service requirements.

6 Maintenance

6.1 Calibration and Adjustment

Information on calibration procedures, frequency, and service interval notifications.

9 Adjustable Parameters

Operation Parameters

Adjustable settings for test results display, breath test volume, and restart period.

Dräger Interlock 5000 Specifications

General IconGeneral
Data LoggingYes
Device TypeAlcohol Tester
TechnologyElectrochemical
Measurement PrincipleElectrochemical
Response Time< 5 seconds
Humidity Range10 to 95 %RH, non-condensing
Power SupplyRechargeable battery
DisplayLCD
AlarmAudible and visual