Reprocessing instructions | Perseus A500 11
Reprocessing
CLIC absorber (disposable absorber)
Observe the instructions for use for the CLIC absorber.
1. Press the release button (A). The CLIC adapter flips open.
2. Pull the CLIC absorber (C) out of the holder (B) in the direction of the arrow.
3. Fold the holder back again until it engages audibly.
4. Dispose of the CLIC absorber in accordance with the instructions for use.
3.4.2.2 Removing the flexible breathing bag arm (option)
1. Unfasten the knurled screws (A) on the attachment piece for the breathing bag
arm.
2. Remove the attachment piece (B) with the breathing bag arm from the socket
(C) on the breathing system. Remove the breathing bag arm.
3901149668
C
B
A